Detail of Recall

Type of Product:Device
Medsafe Reference:23139
Brand Name:CN - MicroTargeting Drive
Model:Model FC1006: microTargeting™ Drive System, Model FC8001: microTargeting™ STar™ Drive System (Manual)
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:FHC Inc
Issue:The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperatively or in a follow-up procedure.
Recall action type:Recall
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/6/2018