Type of Product:Device
Medsafe Reference:23114
Brand Name:RaySearch_Radiation therapy treatment planning system
Model:RayStation 2.5, RayStation 3.5, RayStation 4.0, RayStation 4.3 (InverseArc 1.0), RayStation 4.5, RayStation 4.7, RayStation 4.9 (RayPlan 1), RayStation 5, RayStation 6, RayPlan 2) and RayStation 7 (RayPlan 7)
Affected:
Software version:
Recalling Organisation:CARSL Consulting
30 Fulford Road
Havelock North 4130
Hastings
Contact Information:09 238 3447
Manufacturer:RaySearch Laboratories AB (publ)
Issue:This is a reminder to follow the instructions listed in the product documentation when creating a photon beam model in RayStation/RayPlan. It is the user's responsibility to validate the beam model for all clinically relevant fields before the system is used to create clinical treatment plans.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/6/2018