| Type of Product: | Device |
| Medsafe Reference: | 23114 |
| Brand Name: | RaySearch_Radiation therapy treatment planning system |
| Model: | RayStation 2.5, RayStation 3.5, RayStation 4.0, RayStation 4.3 (InverseArc 1.0), RayStation 4.5, RayStation 4.7, RayStation 4.9 (RayPlan 1), RayStation 5, RayStation 6, RayPlan 2) and RayStation 7 (RayPlan 7) |
| Affected: | |
| Software version: | |
| Recalling Organisation: | CARSL Consulting 30 Fulford Road Havelock North 4130 Hastings |
| Contact Information: | 09 238 3447 |
| Manufacturer: | RaySearch Laboratories AB (publ) |
| Issue: | This is a reminder to follow the instructions listed in the product documentation when creating a photon beam model in RayStation/RayPlan. It is the user's responsibility to validate the beam model for all clinically relevant fields before the system is used to create clinical treatment plans. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 11/6/2018 |