Detail of Recall

Type of Product:Device
Medsafe Reference:23105
Brand Name:qUAntify Control
Model:
Affected:qUAntify Control - Lots 78770, 78780, qUAntify Plus Control - Lots 80510, 80520, 80530, 80540, 80550, 80560, qUAntify Plus Control - Lots 79610, 79620, 79630, 79640, 79650, 79660
Software version:
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Bio-Rad Laboratories Inc
Issue:The manufacturer has observed that the values for Blood and Ketones may gradually decrease and values for pH may gradually increase over the shelf life of certain product lots. Information has been added to the Storage and Stability section in the package insert.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/6/2018