Detail of Recall

Type of Product:Device
Medsafe Reference:23103
Brand Name: Caesarea Syringe Pump
Model: NIKI T34 Syringe Driver pumps
Affected:
Software version:
Recalling Organisation:Paragon Care Group New Zealand Limited
69 Elizabeth Knox Place
Glen Innes
AUCKLAND
Contact Information:0508 654 258
Manufacturer:Caesarea Medical Electronics Ltd (Israel)
Issue:It has been identified that there can be a 2mm +/- overall length difference between various manufacturers PP3 batteries that could be used in NIKI T34 Ambulatory Syringe Pumps.
This could result in movement within the battery housing leading to a possible loss of connection, resulting in the pump powering down and a possible under infusion of pain medication occurring.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/6/2018