Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:23090
Brand Name:Stryker Leibinger Craniofacial fixation plate, non-biodegradable
Model:
Affected:Multiple Lot numbers
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker Leibinger GmbH & Co KG
Issue:The current packaging and IFU indicate a five-year shelf life. During long-term product testing, the manufacturer discovered that after a three-year term, there is a potential for the external packaging (sterile barrier) to become compromised. Such compromised packaging results in the loss of the vacuum seal allowing for contamination of the inside peel pouch.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/6/2018
Hide menus