Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:23089
Brand Name:Medtronic Xomed Burs
Model:BLADE 1883504 TRICUT 5PK 3.5MM, BLADE 1883516 RAD60 3PK PEDIATRIC 3.5MM
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Xomed
Issue:The manufacturer has indentified that the irrigation pin used to manufacture (molding) the hub outer tube was incorrectly placed in the mold which resulted in blockage of the irrigation lumen.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/6/2018