Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 23011 |
| Brand Name: | ABL90Flex analyser |
| Model: | All analysers with software version 3.3MR1 or lower |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Radiometer Pacific Suite 1 1/30 Benedicts Street Eden Terrace Auckland 1010 |
| Contact Information: | 0800 723 722 |
| Manufacturer: | Radiometer Medical ApS |
| Issue: | The manufacturer has recently become aware that the ABL90 FLEX in a very rare specific situation can aspirate Cal 2 solution instead of rinse solution. The specific situation can occur when the following take place: 1. The analyzer is in ready mode 2. Operator lifts the inlet and then closes it again, thereby initiating a rinse 3. Immediately after lifts the inlet and quickly closes it again |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 23/5/2018 |