Detail of Recall

Type of Product:Device
Medsafe Reference:23011
Brand Name:ABL90Flex analyser
Model:All analysers with software version 3.3MR1 or lower
Affected:
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:The manufacturer has recently become aware that the ABL90 FLEX in a very rare specific situation can aspirate Cal 2 solution instead of rinse solution.
The specific situation can occur when the following take place:
1. The analyzer is in ready mode
2. Operator lifts the inlet and then closes it again, thereby initiating a rinse
3. Immediately after lifts the inlet and quickly closes it again
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/5/2018