Detail of Recall

Type of Product:Device
Medsafe Reference:22996
Brand Name:Philips OmniDiagnostic
Model:OmniDiagnost Eleva and OmniDiagnost Classic
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The fixation of the upper and lower tilt actuator, of the Omnidiagnost system might break off and the table will start to rotate from 0 to + 90 /- 90 degree with high speed. This rotating movement can not be stopped by the user.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/5/2018