Detail of Recall

Type of Product:Device
Medsafe Reference:22989
Brand Name:Medical Intelligence HexaPOD evo RT System
Model:HexaPOD evo Module.
Affected:
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Medical Intelligence Medizintechnik GmbH
Issue:It has been noted that upon completion of a 6D workflow, the HexaPOD evo Module may be in a tilted state, if it is moved to the position used for 3D (non-iGUIDE) treatments. In a subsequent 3D workflow, a possible incorrect position of the patient, due to a tilt of the HexaPOD, may be difficult to detect without verification (e.g. imaging). This description is not relevant for an iGUIDE workflow (6D). The problem occurs only when you switch from a 6D to a 3D workflow without iGUIDE.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/5/2018