Type of Product:Device
Medsafe Reference:22974
Brand Name:Liquichek Urine Chemistry Control Level 2
Model:
Affected:Lot Number 68512, 68510
Software version:
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Bio-Rad Laboratories Inc
Issue:The manufacturer has received multiple customer complaints indicating some analyte values in certain vials of Liquichek Urine Chemistry Control Level 2, Lot 68512, are recovering outside the package insert ranges with high CVs.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/5/2018