Detail of Recall

Type of Product:Device
Medsafe Reference:22973
Brand Name:Computed Tomography System - Yokogawa
Model:CT660, CT670, CT680, CT540 systems
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Yokogawa Medical Systems Ltd
Issue:The manufacturer has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry. This issue does not affect patients or operators external to the equipment.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/5/2018