| Brand Name: | AKRON and Omega Couch Ranges |
| Model: | |
| Affected: | Serial numbers within a range between P0167499 and P0418644 |
| Software version: | |
| Recalling Organisation: | Arjo New Zealand Limited 34 Vestey Drive Mt Wellington AUCKLAND 1060 |
| Contact Information: | 0800 000 151 |
| Manufacturer: | ArjoHuntleigh AB |
| Issue: | The manufacturer has received a complaint of a Akron Tilt Table middle section frame breakage. The investigation showed that one of critical welds in the part was missing. Please note that not all devices are affected by this defect. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 10/5/2018 |