Detail of Recall

Type of Product:Device
Medsafe Reference:22960
Brand Name:AKRON and Omega Couch Ranges
Model:
Affected:Serial numbers within a range between P0167499 and P0418644
Software version:
Recalling Organisation:Arjo New Zealand Limited
34 Vestey Drive
Mt Wellington
AUCKLAND 1060
Contact Information:0800 000 151
Manufacturer:ArjoHuntleigh AB
Issue:The manufacturer has received a complaint of a Akron Tilt Table middle section frame breakage. The investigation showed that one of critical welds in the part was missing. Please note that not all devices are affected by this defect.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/5/2018