Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:22958
Brand Name:Askina Gel
Model:
Affected:ASKINA GEL 15G (Item Code: 001419S),
Software version:
Recalling Organisation:B Braun New Zealand Ltd
HSBC Tower
188 Quay Street
Parnell
AUCKLAND 1010
Contact Information:021 784 724
Manufacturer:B Braun Hospicare Ltd
Brand Name:Askina Calgitrol Paste
Model:
Affected:ASKINA CALGITROL PASTE (Item Code 6241510)
Software version:
Recalling Organisation:B Braun New Zealand Ltd
HSBC Tower
188 Quay Street
Parnell
AUCKLAND 1010
Contact Information:021 784 724
Manufacturer:B Braun Hospicare Ltd
Issue:The manufacturer has determined that the irradiation dose qualified for sterilization of the Askina Gel 15G and Askina Calgitrol Paste was too low. The germ reduction through gamma irradiation may not have reached the requested sterility assurance level of 1x10-6.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/5/2018