Detail of Recall

Type of Product:Device
Medsafe Reference:22951
Brand Name:Heartware - Ventricular Assist System (HVAD)
Model:
Affected:Controller/ Controller Kits, DC Adapter, AC Adapter, Batter Pack
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:HeartWare Inc
Issue:It was identified that a transient interruption of the electrical connection between the HVAD System power source and the HVAD Controller can occur under 2 configurations that may result in unintended power switching to the secondary power source and/or unexpected audible tones.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Healthcare Professional
Date Commenced:8/5/2018