Detail of Recall

Type of Product:Device
Medsafe Reference:22949
Brand Name:Artis Q, Q.zen, zeego, zee systems
Model:Catalogue number – 10094137, 10094135, 10280959, 10848281, 10848283, 10094141
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Issue:The manufacturer has indentified that the Large Display does not show an image due to a technical problem. This might occur after the Large Display returns from power save mode. The Large Display stays black without showing an error message. X-ray is still possible. The problem occurs sporadically and only when the display is returning from power save mode. It does not occur during an ongoing procedure.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/5/2018