Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:22943
Brand Name:Analgesic Gas Blender
Model: Equinox Advantage
Affected:
Software version:
Recalling Organisation:BOC Limited
Lower Hutt Gas and Gear Centre
Hutt Park Road
Gracefield
LOWER HUTT
Contact Information:
Manufacturer:O-Two Medical Technologies Inc
Issue:A recent incident with the Equinox Advantage led to the oxygen supply to the patient being cut off. The manufacturer has determined that the fault was caused by a disengaged piston seat in the equalizer that has now been redesigned, to ensure this problem does not reoccur.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/5/2018