Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 22943 |
| Brand Name: | Analgesic Gas Blender |
| Model: | Equinox Advantage |
| Affected: | |
| Software version: | |
| Recalling Organisation: | BOC Limited Lower Hutt Gas and Gear Centre Hutt Park Road Gracefield LOWER HUTT |
| Contact Information: | |
| Manufacturer: | O-Two Medical Technologies Inc |
| Issue: | A recent incident with the Equinox Advantage led to the oxygen supply to the patient being cut off. The manufacturer has determined that the fault was caused by a disengaged piston seat in the equalizer that has now been redesigned, to ensure this problem does not reoccur. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 8/5/2018 |