Detail of Recall

Type of Product:Device
Medsafe Reference:22936
Brand Name:CellaVision DI-60 version 6.0.1 and 6.0.2
Model:All on sw version 6.0.1 and 6.0.2.
Affected:
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Sysmex Corporation
Issue:A bug has been discovered in versions 6.0.1 and 6.0.2 of CellaVision DI-60 software where some comments are not sent to LIS.
The problem occurs only when a specific workflow involving multi-slide is selected as described below:
a)Multiple slide workflow is selected (i.e. > 1 slide is analyzed on a sample no.), AND
b)Operator adds a comment on the last slide in the WBC, RBC or Plt comment boxes in the DI-60 sw after it has been signed/reported.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/5/2018