Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 22936 |
| Brand Name: | CellaVision DI-60 version 6.0.1 and 6.0.2 |
| Model: | All on sw version 6.0.1 and 6.0.2. |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Roche Diagnostics (NZ) Ltd 15 Rakino Way Mt Wellington AUCKLAND 1644 |
| Contact Information: | 0800 363 622 |
| Manufacturer: | Sysmex Corporation |
| Issue: | A bug has been discovered in versions 6.0.1 and 6.0.2 of CellaVision DI-60 software where some comments are not sent to LIS. The problem occurs only when a specific workflow involving multi-slide is selected as described below: a)Multiple slide workflow is selected (i.e. > 1 slide is analyzed on a sample no.), AND b)Operator adds a comment on the last slide in the WBC, RBC or Plt comment boxes in the DI-60 sw after it has been signed/reported. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 7/5/2018 |