Detail of Recall

Type of Product:Device
Medsafe Reference:22934
Brand Name:Dimension clinical chemistry system and Dimension Vista System
Model:Gentamicin (GENT) Flex reagent cartridge
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
55 Hugo Johnston Drive
Penrose
Auckland
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has confirmed Dimension Gentamicin (GENT) lot BB8159 may exhibit inaccuracy for Quality Control (QC) and patient samples at the low end of the Analytical Measurement Range. Maximum negative bias of 100% and 48% were observed for patient samples at 0.9 mg/L [1.94 µmol/L] and 1.9 mg/L [4.1 µmol/L], respectively. Patient samples at 2.6 mg/L [5.62 µmol/L] did not show a bias. The bias is due to the calibration curve not showing separation between the Level 1 and Level 2 calibrators.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/5/2018