Detail of Recall

Type of Product:Device
Medsafe Reference:22901
Brand Name:Therapeutic plasma exchange system
Model:955052, 114870, 114489, 107493, 113082 and 113874
Affected:
Software version:
Recalling Organisation:Baxter Healthcare Ltd
33 Vestey Drive
Mt Wellington
AUCKLAND
Contact Information:0800 229 837
Manufacturer:Gambro Lundia AB
Issue:Manufacturer has identified that all Prismaflex Control Units exhibit a failure mode with the pump module electronics. The failure mode may result in a Voltage Out of Range malfunction alarm, which causes the device to enter a safe state and become inoperable until it is serviced. Manufacturer will be releasing new firmware that will prevent the malfunction from occurring
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/5/2018