Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 22860 |
| Brand Name: | GE Treadmill |
| Model: | T2100 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | Full Vision Inc |
| Issue: | A performance issue with customer owned spare parts, T2100 Microflex drive (2026182-002 or 2026182-004), was not addressed with a previous safety correction in 2015. If a spare part drive does not have the FMI correction label attached and is GE part number (2026182-002 or 2026182-004), or ABB part number (FMH2A09TR-EN43EQ, FMH2A09TR-EN43E, FMH2A09TR-EN43E/12 or FMH2A09TR-EN43GZ), the firmware on the drive was not updated and spare part drive has not been corrected. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 26/4/2018 |