Type of Product:Device
Medsafe Reference:22836
Brand Name:Philips Intera and Achieva MRi Systems
Model:Intera 1.5T and Achieva 1.5T.
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:In rare cases, helium gas may escape into the examination room when superconductivity is lost (also known as quenching of the magnet) and at the same time the designed helium ventilation path is impeded.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/4/2018