Detail of Recall

Type of Product:Device
Medsafe Reference:22831
Brand Name:Philips MultiDiagnost Eleva
Model:MultiDiagnost Eleva 708032, MultiDiagnost Eleva with Flat Detector 708034, MultiDiagnost Eleva with Flat Detector 708038
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has become aware that the fixation of the upper and lower tilt actuator of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 / - 90 degrees with high speed if the point of gravity of the table is beyond its rotation point. This rotating movement can not be stopped by the user, which potentially could lead to the patient falling off the table.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/4/2018