Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 22820 |
| Brand Name: | Sysmex CV-50 (conveyor on XN-9000 systems) |
| Model: | SN: 16489 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Roche Diagnostics (NZ) Ltd 15 Rakino Way Mt Wellington AUCKLAND 1644 |
| Contact Information: | 0800 363 622 |
| Manufacturer: | Sysmex Corporation |
| Issue: | There is a possibility that the CV-50 power supply board may become damaged if a fluid leak from the XN traveled down to the power supply unit of the CV-50. This could cause a short-circuit on the elements of the power supply board which may result in burning of the PCB board. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 17/4/2018 |