Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 22799 |
| Brand Name: | Stryker Medtech - I - 47675; Ultrasonic surgical system torque wrench, reusable |
| Model: | |
| Affected: | Torque Wrench 90 Degree, Universal |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Stryker Medtech K.K. |
| Brand Name: | Stryker Medtech - IIb - 44750; Soft-tissue ultrasonic surgical system handpiece, foot controlled reusable |
| Model: | |
| Affected: | Small Long Angled Handpiece, Hyper Small Angled Handpiece |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Stryker Medtech K.K. |
| Brand Name: | Stryker Medtech - IIb - 36273; Ultrasonic surgical system generator |
| Model: | Sonopet 110V Console with Foot Pedal and Pole, Sonopet 100V Console with Foot Pedal and Pole, Sonopet 230V Console with Foot Pedal and Pole |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Stryker Medtech K.K. |
| Issue: | Additional information is being added to the Instructions For Use (IFU) for Sonopet® Ultrasonic Products. Ultrasonic surgical aspirator devices use an oscillating tip to cause tissue fragmentation through the delivery of ultrasound energy to target tissue. The potential for tissue dissemination is mitigated, but not completely eliminated, by the use of suction/aspiration. Although FDA is not aware of reports of dissemination or upstaging of occult uterine malignancies related to ultrasonic surgical aspirators at this time, the agency is recommending that all device manufacturers add a contraindication against use for the removal of uterine fibroids. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 11/4/2018 |