| Type of Product: | Device |
| Medsafe Reference: | 22782 |
| Brand Name: | RaySearch_Radiation therapy treatment planning system |
| Model: | RayStation 6 (RayPlan 2), RayStation 7 (RayPlan 7) |
| Affected: | |
| Software version: | |
| Recalling Organisation: | CARSL Consulting 30 Fulford Road Havelock North 4130 Hastings |
| Contact Information: | 09 238 3447 |
| Manufacturer: | RaySearch Laboratories AB (publ) |
| Issue: | The manufacturer has indentified an issue with the photon Collapsed Cone dose calculation in RayStation 6 (RayPlan 2) and RayStation 7 (RayPlan 7). For machines with fixed jaws (e.g., Elekta BM and Vero) and machines with the MLC closer to the source than both x- and y-jaws (e.g., Elekta Synergy with MLCi/MLCi2), the dose calculation accuracy may in some situations be less than expected. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 6/4/2018 |