Detail of Recall

Type of Product:Device
Medsafe Reference:22782
Brand Name:RaySearch_Radiation therapy treatment planning system
Model:RayStation 6 (RayPlan 2), RayStation 7 (RayPlan 7)
Affected:
Software version:
Recalling Organisation:CARSL Consulting
30 Fulford Road
Havelock North 4130
Hastings
Contact Information:09 238 3447
Manufacturer:RaySearch Laboratories AB (publ)
Issue:The manufacturer has indentified an issue with the photon Collapsed Cone dose calculation in RayStation 6 (RayPlan 2) and RayStation 7 (RayPlan 7). For machines with fixed jaws (e.g., Elekta BM and Vero) and machines with the MLC closer to the source than both x- and y-jaws (e.g., Elekta Synergy with MLCi/MLCi2), the dose calculation accuracy may in some situations be less than expected.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/4/2018