Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 22778 |
| Brand Name: | X-ray System |
| Model: | C-Arm X-Ray System |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Ortho Medics Limited 586 Great South Road Ellerslie Auckland 1051 |
| Contact Information: | 09 527 1203 |
| Manufacturer: | Ortho Scan Inc |
| Issue: | The manufacturer has determined that there is a potential Hazard where the user should never select the Digital Zoom when Fluoroscopic X-ray beam is active. This allows the system to expose a larger x-ray field exposure than allowed beyond the image receptor field of view displayed on the monitor of the system by > 3% of SID. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 6/4/2018 |