Detail of Recall

Type of Product:Device
Medsafe Reference:22777
Brand Name:AdVIA Chemistry Urea Nitrogen
Model:
Affected:kit lots 408143 and 418290 for use on ADVIA® 1800, 2400 and XPT Chemistry Systems
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has confirmed that the Urea Nitrogen reagent kit lots 408143 and 418290 for use on ADVIA 1800, 2400 and XPT Chemistry Systems demonstrate an increased incidence of Calibration Failures. Results cannot be generated by the system when calibration fails.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/4/2018