Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:22751
Brand Name:Mako - IIa - 44054 - RIO
Model:
Affected:Multiple lot numbers
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Mako Surgical Corp
Issue:The manufacturer has received a report that a hinge pin disassociated from the Mako Onlay Insert Extractor. The disassociated hinge pin was discovered on the back table prior to being used during surgery and a backup device was available and used to complete the procedure.
The manufacturer has expanded the scope of this lot-specific recall. The following lots have been added to the scope: 19150517, 19160517.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/12/2018