Detail of Recall

Type of Product:Device
Medsafe Reference:22739
Brand Name:HemoCue White Blood cell analysers
Model:
Affected:All serial numbers
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Hemocue AB
Brand Name:HemoCue Microcuvettes
Model:
Affected:All lots
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Hemocue AB
Issue:The manufacturer has become aware that there is a need to clarify that use of the HemoCue WBC Diff system on patients on chemotherapy, other immunosuppressive therapies or on drugs with immunosuppressive effects, has not been fully evaluated by HemoCue AB. If incorrect results are obtained for these patient populations it could potentially lead to a delay in a dose modification and increased infection risk.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/3/2018