Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:22736
Brand Name:Lights - Ardon
Model:VOLISTA 400 and VOLISTA 600 P/N ARD568801560
Affected:
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Maquet SA
Issue:The manufacturer has indentified a failure of the control keypad. This can result in the cupola not turning on or off. The manufacturer has received numerous complaints where this problem happened on 3 different keypad versions. The investigation which, included evaluation of worst-case scenarios, concluded that there is only a remote chance that the problem may cause injury.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/3/2018