Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:22722
Brand Name:Cook Stonebreaker Pneumatic Lithotripter
Model:
Affected:All lots
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:Cook Incorporated
Brand Name:Stonebreaker CO2 Cartridge
Model:
Affected:All lots
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:Cook Incorporated
Brand Name:Stonebreaker Single Use Probe
Model:
Affected:All lots
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:Cook Incorporated
Issue:The manufacturer has identified that the interior of the product may not be sterilised to the appropriate sterility assurance level when following the reprocessing instructions provided in the Instructions for Use (IFU). Potential adverse events that may occur include urinary tract infection (UTI), pyelonephritis, and urosepsis.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/3/2018
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /