Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:22719
Brand Name:Sequoia Cypress
Model:ACUSON SC2000 ultrasound with 5.0 software (VB20A, VB20B, or VB20C) and the ECG (physio) hardware and Auxiliary (AUX IN/OUT) cable
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Medical Solutions USA Inc
Issue:The manufacturer has discovered that in some cases the ECG signal may flatline due to electromagnetic interference during the use of electrosurgical equipment. As a result a potential risk in delay of the procedure may occur.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/3/2018