Detail of Recall

Type of Product:Device
Medsafe Reference:22718
Brand Name:Gambro Filter for ultrafiltration of water for intravenous
Model:
Affected:Product code 112062
Software version:
Recalling Organisation:Baxter Healthcare Ltd
33 Vestey Drive
Mt Wellington
AUCKLAND
Contact Information:0800 229 837
Manufacturer:Gambro Dialysatoren GmbH & Co KG
Issue:On March 28, 2018, the manufacturer issued an Urgent Device Correction Communication for the Ultrafilter U9000 due to reports of leaks during regular clinical use. In order to reduce the patient risk associated with the occurrence of leaks when the Ultrafilter is used in conjunction with AK96 and AK98 v1 dialysis machines, an update to the Instructions For Use (IFU) is in progress which will reduce the maximum lifetime usage to 60 days, or 100 disinfection cycles, with a maximum of 8 sodium carbonate and 12 sodium hypochlorite disinfection cycles.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:25/3/2019