Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 22712 |
| Brand Name: | Cleveland CT |
| Model: | Brilliance Big Bore(Radiology), Brilliance Big Bore (Oncology), Brilliance iCT, Brilliance iCT SP, Ingenuity Core, Ingenuity Core128, Ingenuity CT, Ingenuity Flex |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems |
| Issue: | Upon closing the Emergency Stop (E-STOP) the system performs an internal check. As a result of the motion controller in the couch, there is a delay in the response to the internal check. This delay may cause the couch carbon top to enter a free float state, which may allow it to move during patient load or unload activities. This free float motion results in the ESTOP being opened due to the unexpected couch motion. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 13/3/2018 |