Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:22712
Brand Name:Cleveland CT
Model:Brilliance Big Bore(Radiology), Brilliance Big Bore (Oncology), Brilliance iCT, Brilliance iCT SP, Ingenuity Core, Ingenuity Core128, Ingenuity CT, Ingenuity Flex
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:Upon closing the Emergency Stop (E-STOP) the system performs an internal check. As a result of the motion controller in the couch, there is a delay in the response to the internal check. This delay may cause the couch carbon top to enter a free float state, which may allow it to move during patient load or unload activities. This free float motion results in the ESTOP being opened due to the unexpected couch motion.

Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/3/2018