Type of Product:Device
Medsafe Reference:22696
Brand Name:Genesee Retractors
Model:
Affected:GEN-SRCL-1 All lots, GEN-SRCL-AR All lots, GEN-SRLP-1 All lots, GEN-CSR-1 All lots
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:Genesee Biomedical Inc
Issue:Corrosion was found to have caused failure in a retractor blade. In the area of the blade where the bearing/button meets the blade body, corrosion can form, weakening the welded joint and may cause the bearing/button to break.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/3/2018