Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:22695
Brand Name:EES Generator 36273
Model:All serial number are affected
Affected:
Software version:
Recalling Organisation:Johnson & Johnson (New Zealand) Limited
507 Mt Wellington Highway
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Ethicon Endo-Surgery Inc
Issue:The manufacturer has confirmed a cybersecurity software vulnerability on Ethicon Generators Gen11 and is issuing a field cybersecurity routine update and patch to address that cybersecurity software vulnerability. This cybersecurity software vulnerability can exist when using the Gen11 system with non- Ethicon disposable devices as it is possible for these devices to bypass the Gen11 security authentication.
After the update of the Gen11 to address the cybersecurity software vulnerability, non-Ethicon manufactured products may no longer work on the Gen11.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/3/2018