Detail of Recall

Type of Product:Device
Medsafe Reference:22694
Brand Name:Neonatal Ventilator
Model:Serial numbers 1240 - 1455
Affected:
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Maquet Critical Care AB
Brand Name:Ventilator
Model:Serial numbers 22611 - 24876
Affected:
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Maquet Critical Care AB
Issue:The manufacturer has identified that under certain conditions, an internal communication error might prevent the safety valve to close automatically. As a consequence, ventilation will stop and a high priority alarm, as well as clinical alarms, will be triggered. We are aware of two instances where this issue has caused desaturation but we have no reports of any permanent injuries.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/3/2018