| Type of Product: | Device |
| Medsafe Reference: | 22694 |
| Brand Name: | Neonatal Ventilator |
| Model: | Serial numbers 1240 - 1455 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Getinge Australia Pty Ltd (NZ) Unit D, Ground Floor, Building A, Millenium Centre 600 Great South Rd Ellerslie Auckland 1051 |
| Contact Information: | 09 272 9039 |
| Manufacturer: | Maquet Critical Care AB |
| Brand Name: | Ventilator |
| Model: | Serial numbers 22611 - 24876 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Getinge Australia Pty Ltd (NZ) Unit D, Ground Floor, Building A, Millenium Centre 600 Great South Rd Ellerslie Auckland 1051 |
| Contact Information: | 09 272 9039 |
| Manufacturer: | Maquet Critical Care AB |
| Issue: | The manufacturer has identified that under certain conditions, an internal communication error might prevent the safety valve to close automatically. As a consequence, ventilation will stop and a high priority alarm, as well as clinical alarms, will be triggered. We are aware of two instances where this issue has caused desaturation but we have no reports of any permanent injuries. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 6/3/2018 |