Type of Product:Device
Medsafe Reference:22664
Brand Name:Syringe Pump
Model:Model 3500 Product Codes 3500-0600-50, 3500-0600-51
Affected:
Software version:
Recalling Organisation:Smiths Medical Australasia Pty Ltd
c/o KPMG, KPMG Centre
18 Viaduct Harbour Avenue
Auckland
Contact Information:0800 444 200
Manufacturer:Smiths Medical ASD Inc
Issue:The manufacturer has identified that certain Medfusion® Syringe Pump Model Series 3500 pumps were assembled with a barrel clamp guide that contained a ridge in the component. This ridge could potentially lead to spring slippage, resulting in the inability of the pump to recognise a syringe or the pump may misidentify the size of syringe loaded.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/2/2018