| Type of Product: | Device |
| Medsafe Reference: | 22664 |
| Brand Name: | Syringe Pump |
| Model: | Model 3500 Product Codes 3500-0600-50, 3500-0600-51 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Smiths Medical Australasia Pty Ltd c/o KPMG, KPMG Centre 18 Viaduct Harbour Avenue Auckland |
| Contact Information: | 0800 444 200 |
| Manufacturer: | Smiths Medical ASD Inc |
| Issue: | The manufacturer has identified that certain Medfusion® Syringe Pump Model Series 3500 pumps were assembled with a barrel clamp guide that contained a ridge in the component. This ridge could potentially lead to spring slippage, resulting in the inability of the pump to recognise a syringe or the pump may misidentify the size of syringe loaded. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 27/2/2018 |