Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 22643 |
| Brand Name: | Philips DynaCAD 3 |
| Model: | DynaCAD 3.4 and 3.5 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | iCAD Inc |
| Issue: | The manufacturer has become aware of a problem which causes images processed with Pharmacokinetic (PK) results and/or DynaCAD calculated Apparent Diffusion Coefficient (ADC) maps have incorrect color overlay results as well as in the case of the ADC images, incorrect values reported. The defect exists on any DynaCAD client system running software versions 3.4 or 3.5. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 21/2/2018 |