Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 22629 |
| Brand Name: | Infinity Acute Care System Monitoring - M540 |
| Model: | All released software versions (from VG2.2 to VG6.0) |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Draeger New Zealand Ltd Unit 4 24 Bishop Dunn Place East Tamaki Auckland 2013 |
| Contact Information: | 0800 559 186 |
| Manufacturer: | Draegerwerk AG & Co KGaA |
| Brand Name: | INFINITY Patient Monitors & Accessories |
| Model: | All released software versions (from VG2.2 to VG6.0); |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Draeger New Zealand Ltd Unit 4 24 Bishop Dunn Place East Tamaki Auckland 2013 |
| Contact Information: | 0800 559 186 |
| Manufacturer: | Draeger Medical Systems Inc |
| Issue: | The manufacturer has determined that when the Infinity M540 patient monitor is placed in Standby mode and the user simultaneously connects the Infinity MCable- Dual Hemo or Infinity MPod - Quad Hemo to the Infinity M540 patient monitor and then proceeds to zero the invasive pressure channel, in rare cases the invasive pressure visual and audible alarms are not activated. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 19/2/2018 |