Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 22618 |
| Brand Name: | VIDAS® and mini VIDAS® systems |
| Model: | |
| Affected: | Ref codes Mimi Vidas 99174, 99733, 99734, 99737, 99739,410416, W1421 and Vidas 93296, 93297, 99011, 99444, 99735,W3205,410417 |
| Software version: | |
| Recalling Organisation: | Medi'Ray New Zealand Ltd 53-55 Paul Matthews Road Albany Auckland 0632 |
| Contact Information: | 0800 633 472 |
| Manufacturer: | Biomerieux SA |
| Issue: | The manufacturer has identified pump clogging and displacement of the color-coded SPR® label (DOT). These pump clogging and DOT displacement could contribute to erroneous results for any assay performed on the VIDAS® and mini VIDAS® system due to a decrease of pipetted volume. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 15/2/2018 |