Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:22557
Brand Name:Clamps
Model:
Affected:STRAUSS PENIS CLAMP 130MM - EF401R
Software version:
Recalling Organisation:B Braun New Zealand Ltd
HSBC Tower
188 Quay Street
Parnell
AUCKLAND 1010
Contact Information:021 784 724
Manufacturer:Aesculap AG
Issue:It has been determined that the plastic sleeves of the returned instrument do not meet the valid specification. Instead of the specified material silicone, the material PVC was used for the manufacturing of the sleeves. This could render affected instruments unusable after reprocessing.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/2/2018
Hide menus
Show menus