Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:22556
Brand Name:Lights - Ardon
Model:
Affected:Spring arm part number, ARD567910974, ARD568301798, ARD568301799
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Maquet SA
Issue:The manufacturer has received complaints concerning unintentional drifting of the cupola during surgery due to prematurely worn out brake screws. The observed movement is a slow rotation of the cupola around the axis, potentially shifting the light patch away from the wound.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/2/2018