Detail of Recall

Type of Product:Device
Medsafe Reference:22552
Brand Name:RaySearch_Radiation therapy treatment planning system
Model:RayStation 4.9 (RayPlan 1), RayStation 5, RayStation 6 (RayPlan 2), RayStation 7 (RayPlan 7)
Affected:
Software version:
Recalling Organisation:CARSL Consulting
30 Fulford Road
Havelock North 4130
Hastings
Contact Information:09 238 3447
Manufacturer:RaySearch Laboratories AB (publ)
Issue:An issue found with the Center Beam in Field functionality in RayStation 4.9 (RayPlan 1), RayStation 5, RayStation 6 (RayPlan 2) and RayStation 7 (RayPlan 7). When more than one beam is defined on the same isocenter, Center Beam in Field (CBF) affects all beams in a way that the user may not expect. After using CBF, the resulting fields are clearly displayed in RayStation/RayPlan for review and approval.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/2/2018