Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 22552 |
| Brand Name: | RaySearch_Radiation therapy treatment planning system |
| Model: | RayStation 4.9 (RayPlan 1), RayStation 5, RayStation 6 (RayPlan 2), RayStation 7 (RayPlan 7) |
| Affected: | |
| Software version: | |
| Recalling Organisation: | CARSL Consulting 30 Fulford Road Havelock North 4130 Hastings |
| Contact Information: | 09 238 3447 |
| Manufacturer: | RaySearch Laboratories AB (publ) |
| Issue: | An issue found with the Center Beam in Field functionality in RayStation 4.9 (RayPlan 1), RayStation 5, RayStation 6 (RayPlan 2) and RayStation 7 (RayPlan 7). When more than one beam is defined on the same isocenter, Center Beam in Field (CBF) affects all beams in a way that the user may not expect. After using CBF, the resulting fields are clearly displayed in RayStation/RayPlan for review and approval. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 2/2/2018 |