Detail of Recall

Type of Product:Device
Medsafe Reference:22550
Brand Name:Bair Hugger® Blankets - Temperature Management System
Model:
Affected:Lots numbers R10359, R10360, R10361, R10362, R10363, R10364, R10365, R 10366;
Software version:
Recalling Organisation:3M New Zealand Ltd
250 Archers Road
Glenfield
AUCKLAND 0627
Contact Information:0800 808 182
Manufacturer:Solventum US LLC
Issue:The manufacturer recently became aware through customer feedback that a design change to 3MTM Bair Hugger Full Access Underbody Warming Blanket, Model 63500 can make these blankets more susceptible to unintended occlusion of airflow resulting in limited or partial inflation during use.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/2/2018