| Type of Product: | Device |
| Medsafe Reference: | 22444 |
| Brand Name: | Artis one |
| Model: | The Artis table top carriage |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Shenzhen Magnetic Resonance Ltd |
| Issue: | The Artis table top carriage is mounted on top of the table base and fixed with a bolt. A fixation ring is used to secure the bolt. During the regular filed observation, it was found in other Artis systems, that in a low number of cases the fixation ring is missing or wrongly installed. In extreme cases if the bolt persistently bears the unexpected force from a non-vertical direction, the bolt might slide out of its base and lead to the table top carriage sliding to the floor |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 15/1/2018 |