Detail of Recall

Type of Product:Device
Medsafe Reference:22444
Brand Name:Artis one
Model:The Artis table top carriage
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Shenzhen Magnetic Resonance Ltd
Issue:The Artis table top carriage is mounted on top of the table base and fixed with a bolt. A fixation ring is used to secure the bolt. During the regular filed observation, it was found in other Artis systems, that in a low number of cases the fixation ring is missing or wrongly installed. In extreme cases if the bolt persistently bears the unexpected force from a non-vertical direction, the bolt might slide out of its base and lead to the table top carriage sliding to the floor
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/1/2018