Detail of Recall

Type of Product:Device
Medsafe Reference:22433
Brand Name:i-STAT Alinity Instrument i-STAT Alinity Analyzer Software
Model:with software version OSi05 (JAMS 79A / CLEW D34).
Affected:
Software version:
Recalling Organisation:Abbott Australasia Pty Ltd (NZ)
Abbott Laboratories NZ
4 Pacific Rise, Mt Wellington
Auckland 1060
Contact Information:61 2 93849717
Manufacturer:Abbott Point of Care Canada Ltd
Issue:Through a combination of internal studies and customer reports, the manufacturer has determined that the following may occure:
1. An increased frequency of Quality Check Failure code 90-02 when testing i-STAT cartridges with the i-STAT Alinity software version OSi05 (JAMS 79A / CLEW D34).
2. When testing with i-STAT cartridges that report pH, pCO2 and pO2, an i-STAT Alinity instrument with software version OSi05 (JAMS 79A / CLEW D34) that is customized to correct for user entered patient temperature may report incorrect pH and/or pCO2 results.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory