Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 22432 |
| Brand Name: | Lights - Ardon |
| Model: | 5214 MAQUET VOLISTA StanOP, TRIPOD and ACCESS Cupolas Surgical lights manufactured before July 2016. |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Getinge Australia Pty Ltd (NZ) Unit D, Ground Floor, Building A, Millenium Centre 600 Great South Rd Ellerslie Auckland 1051 |
| Contact Information: | 09 272 9039 |
| Manufacturer: | Maquet SA |
| Issue: | The manufacturer has received complaints indicating that the central handle holder detaches easily from the cupola during use. It was determined that the cause of this malfunction is excessive tightening of the screws during mounting of the handle holder, resulting in the rivets loosening and detaching from the central handle holder ring. |
| Recall action type: | Correction |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 12/1/2018 |