Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 22431 |
| Brand Name: | Patient monitor, central unit |
| Model: | |
| Affected: | CSCS hardware platforms (MAI700 and MAS700) with V2.0.2 software only. |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems Information Technologies Inc |
| Issue: | After boot-up or system start up, CSCS V2.0.2 units may display unexpected NO COMM notifications in the Patient Multi-Viewer or Single Viewer window when the CSCS is connected to both the Mission Critical (MC) and the Information eXchange (IX) networks. If a NO COMM notification is displayed, all patient waveforms, parameter data and alarms are lost at the CSCS and patients admitted to Telemetry transmitters are unmonitored unless monitored via other central stations, bedside monitors or direct clinical observation. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 12/1/2018 |