Detail of Recall

Type of Product:Device
Medsafe Reference:22423
Brand Name:i-STAT Kaolin ACT Cartridges
Model:
Affected:List number 03P87-25
Software version:
Recalling Organisation:Abbott Australasia Pty Ltd (NZ)
Abbott Laboratories NZ
4 Pacific Rise, Mt Wellington
Auckland 1060
Contact Information:61 2 93849717
Manufacturer:Abbott Point of Care Canada Ltd
Issue:To regards to the drug aprotinin (Trasylol), which was previously withdrawn from many markets but may be available in some countries/regions. Internal studies, have identified that the i-STAT Kaolin ACT test may be prolonged immediately after the administration of an initial full dose aprotinin bolus of 400 kallikrein inhibitor units (KIU)/mL. Once the drug has fully distributed throughout the body there is no impact on the ACT result reported. For aprotinin, 200-280 KIU/mL is considered a therapeutic or steady state concentration during CPB.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/1/2018